Key takeaways
- Choose the method by the element that decides the grade: XRF for the main metals, spark OES or light-element XRF for aluminium, OES or combustion for carbon.
- Spark OES works on solid metal at the yard; ICP-OES is a laboratory method on dissolved drillings or composites, and contracts should say which is meant.
- Monel K-500 against 400, aluminium-bearing superalloys, low-carbon grades, rhenium alloys and turnings are the usual cases where XRF alone is not enough.
- Most lots combine XRF on every piece with OES or laboratory analysis on a representative, split sample.
Every nickel alloy lot raises the same practical question before anyone talks about value: what test will both sides accept? A bundle of Hastelloy C-276 plate crops can be settled by handheld XRF in ten minutes. A pallet of Monel bar that might be 400 or K-500 cannot. A drum of mixed superalloy revert headed for a vacuum melter needs a laboratory.
The XRF vs OES choice comes down to one idea: choose the method by the element that decides the grade. This guide compares handheld XRF, spark OES and laboratory analysis (ICP-OES and combustion) element by element and lot by lot, explains why their results sometimes disagree, and shows how buyers and sellers usually combine them.
What Is the Difference Between XRF and OES?
XRF measures X-rays given off by the surface of the metal when it is irradiated, while OES measures light given off by a small amount of metal vaporised by an electric spark. That difference in physics explains almost everything else.
Because XRF only excites atoms and removes nothing, it is non-destructive, needs little preparation and fits in one hand. But light elements give off X-rays too weak to measure well, and carbon gives nothing usable at all. Spark OES does burn the metal: a ground spot is sparked under argon, and the light from the plasma is split into wavelengths, each element having its own lines. Carbon, phosphorus, sulfur and boron emit in the far ultraviolet, which air absorbs, so OES optics run under vacuum or argon. That is how OES reads the elements XRF cannot.
Handheld XRF
- Measures fluorescent X-rays from the surface
- Preparation clean bare spot
- Damage none
- Carbon not measurable
- Aluminium only with a light-element setup
- Equipment handheld, battery powered
Spark OES
- Measures light emitted by a spark under argon
- Preparation ground flat spot
- Damage small burn mark
- Carbon measured
- Aluminium measured
- Equipment transportable unit with an argon supply
For the details of how a handheld analyser reads a nickel alloy, and its specific blind spots, see XRF nickel alloy identification.
XRF vs OES vs Lab Analysis Side by Side
Handheld XRF wins on speed and convenience, spark OES on light elements and carbon, and the laboratory on precision, trace elements and lot-level evidence. The table sets out the practical differences for scrap work.
| Handheld XRF | Spark OES | Laboratory (ICP-OES, combustion) | |
|---|---|---|---|
| Time per result | Seconds per piece | Around a minute per burn, plus grinding | Days, including sample transport |
| What it tests | The spot under the window | The sparked spot | Drillings, chips or a composite of the lot |
| Main metals (Ni, Cr, Fe, Mo, Nb, W, Co, Cu) | Yes | Yes | Yes |
| Aluminium, silicon | Light-element setup only | Yes | Yes |
| Carbon, sulfur | No | Yes | Yes (combustion) |
| Trace elements at ppm levels | No | Limited | Yes, with the right technique |
| Effect on the piece | None | Burn mark on a ground spot | Sample is cut or drilled out |
| Typical role in scrap | Sorting, piece-by-piece checks | Light elements, carbon, close pairs | Lot chemistry, superalloy revert, disputes |
The burn mark and grinding are irrelevant for scrap headed to a furnace. They matter only for material that might be resold for use, such as surplus pipe or bar.
Which Elements Can Each Method Measure?
The element that decides the grade decides the method, so it helps to see coverage element by element.
| Handheld XRF | Light-element XRF | Spark OES | ICP-OES lab | Combustion lab | |
|---|---|---|---|---|---|
| Ni, Cr, Fe, Mo, Co, Cu | ● | ● | ● | ● | ○ |
| Nb, W, Ta | ● | ● | ● | ● | ○ |
| Ti | ● | ● | ● | ● | ○ |
| Al, Si, Mg | ○ | ● | ● | ● | ○ |
| C, S | ○ | ○ | ● | ○ | ● |
| Re, Hf | ◐ | ◐ | ◐ | ● | ○ |
| ppm trace elements | ○ | ○ | ◐ | ● | ○ |
● yes ◐ partly ○ no
Two points stand out. First, nothing in the main-metal rows separates the methods, which is why XRF handles most wrought grades on its own. Second, no single method covers everything: the laboratory pairs ICP-OES for metals with combustion for carbon and sulfur.
Spark OES or ICP-OES: What Is the Difference?
Both measure emitted light, but spark OES vaporises solid metal directly with a spark, while ICP-OES dissolves a sample in acid and sprays it into a hot argon plasma. The shared "OES" causes real confusion in enquiries and contracts.
- Spark OES works on solid metal, on site or at a yard, and gives a result in about a minute. It needs a flat, ground surface and a calibration suited to nickel alloys.
- ICP-OES is a laboratory method. Drillings or chips are weighed, dissolved and analysed against prepared standards. It is slower, but it handles turnings and composites that cannot be sparked, and it reaches lower levels for many elements.
Laboratories also run benchtop wavelength-dispersive XRF, which is far more precise than a handheld unit and is covered for nickel-base alloys by ASTM E2465. So "lab analysis" is not one method: it is whichever combination of ICP-OES, laboratory XRF, wet chemistry and combustion the laboratory uses to report every element the buyer asked for.
When a buyer writes "OES report required", ask which one they mean. A spark OES printout from the loading point and an ICP-OES certificate of analysis from a laboratory are different documents with different weight.
Why Carbon and Sulfur Need Combustion Analysis
Carbon and sulfur are measured in the laboratory by burning a weighed sample in oxygen and measuring the gases produced. ASTM E1019 covers the determination of carbon, sulfur, nitrogen and oxygen in steel, iron, nickel and cobalt alloys by combustion and inert gas fusion techniques.
Why bother on scrap? Carbon separates low-carbon grades from their standard versions, such as 316L from 316, and confirms that Ni-Resist scrap is nickel cast iron rather than a wrought alloy. Sulfur is a concern for melters because it weakens nickel alloys at grain boundaries, and cutting oils and residues on turnings can add it. For the wider family of harmful trace elements, see tramp elements in remelting.
Which Method Should You Use for Each Lot?
Start with XRF for anything wrought and nickel-based, and escalate only when a light element, carbon or a trace element decides the grade or the value.
| Material | Usual method | Why |
|---|---|---|
| Inconel, Hastelloy and Incoloy solids | Handheld XRF | Separating elements (Fe, Mo, Nb, W, Cu, Cr) all read well |
| Monel K-500 against 400 | Light-element XRF or spark OES | Separated by about 2.7% aluminium and 0.6% titanium |
| Inconel 713C, Haynes 214, Nimonic 80A | Light-element XRF or spark OES | Aluminium is part of the grade |
| Cast superalloy revert | XRF to sort, then lab on a sample | Al, Hf, Re and trace elements decide the alloy and its use |
| Rhenium-bearing blades such as René N5 | Laboratory | Rhenium overlaps tungsten and tantalum on XRF |
| Low-carbon against standard grades | Spark OES or combustion | Only carbon differs |
| Ni-Resist castings | Spark OES | Carbon confirms cast iron |
| Titanium Grades 1 and 2 | Paperwork, hardness or lab | They differ mainly in oxygen and iron |
| Turnings of any grade | XRF on several samples, plus a lab composite | One spot reading cannot represent a drum |
- Do aluminium or magnesium decide the grade?YesLight-element XRF or spark OES
- Does carbon decide the grade or matter to the buyer?YesSpark OES or combustion analysis
- Does the alloy carry rhenium, or do ppm trace limits apply?YesLaboratory analysis on a representative sample
- Is the lot turnings, fines or a mixed composite?YesXRF on several samples plus a lab composite
- Handheld XRF on every piece is usually enough
What About Handheld LIBS?
Handheld LIBS (laser-induced breakdown spectroscopy) sits between XRF and spark OES: a laser vaporises a tiny spot and the instrument reads the emitted light. Depending on the model, it can read light elements, including carbon in some units, in a handheld format. The third edition of API RP 578, published in 2018, added LIBS to the methods it discusses for material verification.
LIBS still needs a clean, ground spot and leaves a small mark. In scrap trading it is used much like portable OES: to settle the light-element questions that XRF leaves open.
Why XRF, OES and Lab Results Sometimes Disagree
Different methods test different amounts of metal from different places, so small disagreements are normal and large ones usually point to the sample, not the instrument.
- Surface against bulk. XRF reads the top fraction of a millimetre. Scale, coatings and chromium-depleted layers shift its result; OES grinding and lab drilling go below them.
- Spot against composite. A spot reading describes one piece. A laboratory composite of turnings describes the average of the drum. If the drum mixes 625 and 718 chips, both are "right".
- Segregation in castings. Cast revert can vary from the root of a blade to the tip, so readings at different points differ.
- Calibration. Each instrument is only accurate for alloy types it was calibrated on. A spark OES set up for steels can misreport a cobalt-rich superalloy.
When results differ, look at the separating element first. If both methods agree on it, the grade is settled even if minor elements wander.
What Should a Sales Contract Say About Testing?
A contract should name the method, the sample, the place of testing and whose result governs if the two sides disagree. Most testing disputes on scrap come from a vague line such as "material subject to analysis", agreed before anyone asked which analysis.
- Method. Handheld XRF, spark OES or a named laboratory method, and the elements to be reported. If carbon or aluminium matters, say so.
- Sample. Every piece, a stated number of pieces per bundle, or a composite built a stated way.
- Place and witness. At the seller's premises, at loading, or on arrival, and whether an independent inspector witnesses it.
- Laboratory. For referee analysis, a laboratory both sides accept, often one accredited to ISO/IEC 17025.
- Tolerance and remedy. What happens if the result falls outside the declared grade: a price adjustment, regrading or rejection.
Agreeing these points before shipment costs nothing. Arguing about them after a container has landed costs freight, time and goodwill.
How Buyers and Sellers Combine Methods
Many professional lots use XRF on every piece and OES or laboratory analysis on a representative sample. The XRF record shows each piece is the declared grade; the OES or lab result proves the light elements, carbon and trace elements for the lot.
For laboratory work, the sample matters as much as the instrument:
- 1Choose piecesfrom different points in the lot, not just the top
- 2Cleanremove oil, scale and coatings, avoid welds
- 3Take drillingsor build a dried composite for turnings
- 4Split in threelaboratory, buyer and retained by the seller
- 5Recordlabel, seal and photograph each split
The retained split is what settles a dispute if the buyer's numbers differ from yours. For export lots, buyers often appoint an independent inspection company to witness sampling and testing at loading; buying nickel alloy scrap from India covers how that is usually arranged. The records that tie readings to pieces are covered in positive material identification for scrap lots, and the relevant ASTM methods are listed on the testing and identification standards page.
When you send us an enquiry, include whichever readings or reports you have, XRF, OES or laboratory, and say which method produced them. We'll review your enquiry on that evidence.
Submit a requirement with the grade, form, quantity and delivery location.
Source a GradeFAQ
Is OES more accurate than XRF?
For carbon and light elements such as aluminium and silicon, yes, because XRF cannot read carbon and reads light elements poorly. For the main metals in nickel alloys, both are accurate enough to identify the grade. The useful question is not which is more accurate overall, but which one measures the element that decides your grade.
Can XRF replace laboratory analysis for scrap?
For sorting wrought nickel alloy solids by grade, usually yes. It cannot replace a laboratory when a contract needs certified lot chemistry, when carbon, sulfur or trace elements matter, or when the alloy contains rhenium. Superalloy revert buyers in particular often confirm incoming lots in a laboratory.
Does spark OES damage the material?
It leaves a small burn mark on a ground flat spot. For scrap going to a furnace this does not matter. For surplus pipe, bar or plate that might be resold for use, the ground spot and burn mark may need to be agreed with the buyer or tested in an inconspicuous place.
Can spark OES test turnings?
Not directly in their loose form, because the spark needs a solid, flat surface. Turnings are usually tested by XRF on several compacted samples and by laboratory ICP-OES on a dried composite. The composite should be dried and mixed well before it is split.
What does ICP stand for in ICP-OES?
Inductively coupled plasma. A dissolved sample is sprayed into an argon plasma heated by a radio-frequency coil, and the light emitted by each element is measured. It is a laboratory method, unlike spark OES, which works directly on solid metal.
Which testing standards apply to XRF and OES for nickel alloys?
ASTM E2465 covers laboratory wavelength-dispersive XRF of nickel-base alloys, ASTM E3047 covers spark OES of nickel alloys, ASTM E1019 covers carbon and sulfur by combustion, and ASTM E1476 is a general guide to identifying and sorting metals. See the testing standards page.
How long does laboratory analysis of a scrap sample take?
It depends on the laboratory, the number of elements and the time taken to send the sample. Expect days rather than minutes. That delay is why lots are usually sorted and checked by XRF first, with a laboratory sample taken in parallel for the elements XRF cannot see.